Properties of Strain-Gauge Load Cells Applied for Stator Reaction Measurement

  • Lubomir Miklanek
Keywords: strain-gauge, load cell, hysteresis, dynamometer, torque

Abstract

This contribution deals with experience with the measurement of the stator reaction of the dynamometer using load cells. Originally, the stator reaction had been measured using a balance mechanism. Within the framework of the author’s research, it was necessary to improve the accuracy of torque measurement of the dynamometer. To accomplish this requirement, the old balance mechanism was replaced by a straingauge load cell. Two various types of the load cells were applied in order to arrange the measuring range to the expecting range of values of the torque. During the calibration, hysteresis in voltage output of the both applied load cells was observed. The maximal value of the hysteresis was about 0.5 % of the measured range of both of the load cells. Moreover, a certain temporary-plasticity of both of the load cells occurred.

Author Biography

Lubomir Miklanek

Josef Bozek Research Center, Faculty of Mechanical Engineering, Czech Technical University in Prague, Czech Republic

Published
2009-12-31
How to Cite
Miklanek, L. (2009). Properties of Strain-Gauge Load Cells Applied for Stator Reaction Measurement. Communications - Scientific Letters of the University of Zilina, 11(4), 48-51. Retrieved from http://journals.uniza.sk/index.php/communications/article/view/1017
Section
Articles